Calibration Kits
Reference kits for NIST, NRC, US Pharmacopoeia and European Pharmacopoeia spectrophotometer validation. All standards are NIST traceable and prepared using NIST guidelines. Custom kits are available upon request.
NIST 2031 UV/VIS Photometric Accuracy Kit (10%T, 30%T, 90%T) (200-700nm)
NIST 2031 UV/VIS Photometric Accuracy Kit (10%T, 30%T, 90%T) (200-700nm)
Firefly's upgraded and improved version of the NIST-2031 UV/VIS photometric accuracy kit.
This kits includes the following filters.
Product Specs:
Optical densities: 0.04au (90%), 0.5au (30%), 1.0au (10%)
Wavelength range: 200-700nm. Standard peaks included on certificate of calibration are 250, 280, 340, 360, 400, 465, 500, 546.1, 590 and 635nm.
Tolerances: FUV-0.04 (+/-0.0019 au); FUV-0.5 (+/- 0.0039au); FUV-1 (+/- 0.0042au)
Wavelength range: 200-700nm. Standard certified calibration points included on the certificate of calibration are 250, 280, 340, 360, 400, 465, 500, 546.1, 590, 635nm. Firefly can provide alternative points in the 200-700nm range upon request or full scan data for informational purposes.
External Dimensions: 12.5 x 12.5 x 45 mm
Capabilities:
Photometric accuracy in the UV and VIS range
Linearity in the UV and VIS range
The 2031 Kit Replaces the majority of liquid and solid-state photometric kits:
Potassium Dichromate
VIS Neutral Density Filters
Nicotinic Acid
Included with 2031 Kit:
Certificates of calibration with NIST traceability
NIST 2031 Kit Sample Certificate of Calibration
Customers Who Bought This Also Bought:
Firefly's HF neutral density solid-state filter series tests photometric accuracy in the VIS range (400-700nm).
WAV-7 Didymium glass combined wavelength accuracy (329-875nm) and photometric accuracy (270-340nm) filter.
Firefly's FUV neutral density solid-state filter series tests photometric accuracy in the UV and VIS range from 200-700nm. Our solid-state nano-deposition technology produces unparalleled stability and keeps the filter in spec for years.
Firefly's FUV Dual Series filters allow you to verify photometric accuracy and stray light in the range of 200-700nm.



![firefly_logo_FINAL [Black].jpg](https://images.squarespace-cdn.com/content/v1/5411d5c0e4b02e1c8b27565a/1434491508803-XB4OF7YDY46Z72L5U5AP/firefly_logo_FINAL+%5BBlack%5D.jpg)

